GE Healthcare · CTLightSpeed260006813
LightSpeed 5.X (60kW)
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260006813 Gantry Failed to change to Commanded Position Velocity of 4.0 secs/rev within 15.000 seconds See LD for complete error Host: STC Proc: Axial Error: 260006813 File: amc_exec.c Line: 3326 Function: AXIAL CONTROL : axial control Scan 20/2/0 Type: None/unknown Exception Level: Pri/Most Severe Time: 17:59:36:615 Log Series: 405 Gantry Failed to change to Commanded Position Velocity of 4.0 secs/rev within 15.000 seconds. (AMC state:6 Cmd: 0x17d400b4 CurVel: 4179 (.001 sec/rev)) Retry Rx to verify probable axial drive or Gantry Balance or Encoder Hardware
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277225 Axial Brake assembly ( Holding Brake for Axial Drive ) See Long Description for more info 2277225 Axial Brake assembly ( Holding Brake for Axial Drive ) Part #: 2277225 Part Type: GP CONSUMABLES Part Weight: 2.27 LBR Repairable: No Returnable: No Hazardous Material: N Description: BRAKE ASSY, H2 AX DRIVE MOTOR FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONIC. MS26561 DSM 1/31/
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.