Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18221360046

LightSpeed 4.X - Octane

Symptom

Cradle will not drive in or out. Cradle will not drive in or out

Error message

Error: 182213 Velocity tracking error See Long Description for the full error code Tue Mar 15 15:28:05 2005 Error: 182213 Host: ETC Process: Cradle File: CmdMotion.c 1.33 7/23/96 Line: 92 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:28:05:770 Log Series: 37 HW error: Velocity tracking error. Expected Velocity 500 (.001 mm/sec) Actual Velocity 9382 (.01 mm/sec) Current Position 60046 (.01 mm) Velocity Tolerance 250 (.001 mm/sec) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-264370G1 CRADLE AMP used on Zeus table (1.X-4.X) See long description for details Part #: 46-264370G1 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: CRADLE AMP FRU: Yes Status: Orderable Comments: FLS - B: IF ORDERING 46-264370G1 FOR TROUBLESHOOTING A CRADLE THAT IS EXHIBITING ERRATIC JERKY MOTIONS, ORDER THE 2120791-9 RELAY (24 V POTTER &BRUMFIELD). MSA6521 5/29/01 CRADLE AMP BD 2/8/01- T3202CE DELISTED DUE TO LOW ICV. NO POOL ADDS. MS12464 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required