GE Healthcare · CTLightSpeed74216757963466403
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Taking a long time to transfer exams across the network
Error message
TX-ERR errors greater than 0 on 'netstat -i' report See Long Description for more details TX-ERR errors greater than 0 on 'netstat -i' report Iface MTU Met RX-OK RX-ERR RX-DRP RX-OVR TX-OK TX-ERR TX-DRP TX-OVR Flg eth0 1500 0 742167 0 0 0 579634 66403 0 0
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console - showprods returns 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console - showprods returns 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.