Pular para o conteúdo principal
GE Healthcare · CTHiSpeed230006527

HiSpeed QX/i - Octane

Symptom

Recon self test failure on boot up. Recon self test failure on boot up

Error message

Error: 230006527 Recon Subsystem Reset FAILED See long description for full error Sun Mar 11 10:36:12 2007 Error: 230006527 Host: cbct_sbc Process: reconmgr File: rm_recon_events.c Line: 786 Function: Image Generation : SBC Processing Job Type: Not defined Scan: 0/0/0 Image: 0/0/0 Recon Seq Id: 0 Recon Subsystem Reset FAILED

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2370825 RIP / Motorola Board, 2400 Series Card - replaces all previous part numbers - Oct 14 2004 See long description and attachment 2370825 RIP / Motorola Board, 2400 Series Card, replaces all previous part numbers - 10/14/2004 Part #: 2370825 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 9.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PPC MVME 2400 ASSY, HAWK3 U3D 256MB FRU: Yes Status: Orderable Comments: *** OBSOLETE PART *** This part has been identified as obsolete and no short-term solution has been identified. FE must identify and order a higher or lower level spare part as required. Contact your support engineer, on-line engineer or Parts ID team should you require support. 212012173 10-2-06 2/3/05 (212001261) New RIP-Bd 2370825 only works with LS2002M5 on LS 2X/3X & HS_QXi. LS 4X needs to run LS2002M5.1. To confirm what software version is running in the console, open a shell and type in check_config Even if the system is running LS2002M5, a LFC will probably be required. To determine if a LFC is required, open a shell and type in: showprods (the pipe symbol) grep Helios_MVME enter] if the prompt returns (Helios_MVME24XX_VxWorks_Product) where XX = any number, then no LFC, run pflash. If the prompt returns something else, LFC is required. GOC2/Linux-Pegasus are compatible with RIP-Bd P/N 2370825, (pflash) will need to be run. The following s/w versions support RIP Board 2370825: HS QX/I (IRIX) = 2401246 (101I.2_HSQXIM5), LS 2X (IRIX) 2395178 (2.4.2_H2.4M5); LS 3X (IRIX) = 2395177 (308I.2_H3.1M5), LS 4X (IRIX) = 5116050 (405I.3_H4.0M5.1), HS QX/I (Linux) = 2401245 (103L.4_HSQXIM4), LS 3X (Linux) = 2400882 (308L.5_H3.3M4), LS 4X (Linux) = 5111370 (405L.8_H4.2M4). Can order software via Coakly. Alejandro Cortez
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.