Pular para o conteúdo principal
GE Healthcare · CTLightSpeedDiscovery23000006450003164000

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50003/5/1 Scan Seq Id: 164000 TAXI data violation detected by forward error correction. Interrupt bit: 4 Violation count: 1 Scan aborted.

Applies to

Models
LightSpeed, Discovery
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2318713-2 Fiber Optic cable Part is not a FRU at this time. Order the next higher FRU for your console type See LD for more Fiber Optic cable Part 2318713-2 is not a FRU at this time (10/30/05). Order 2293409 and use this part to get the system up. Part #: 2318713-2 Part Type: Not Available Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: Not Available FRU: No Status: Non Orderable Part #: 2293409 Part Type: GP CONSUMABLES Part Weight: 0.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE FIBER OPT FRU: Yes Status: Orderable Comments: FLS - CABLE FIBER OPT HARNESS ADDED VIA PCN 220275. USED ON WARP3. MS26561 DSM 8/25/01
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.