Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

ECG Trace fails to display on the console. See Long Description for more details. ECG Trace intermittently fails to display on the console, requiring a software reboot to recover. Problem seen after upgrading to 06MW03.4 software. The ECG trace works the first time after a reboot but after any interruption, the trace cannot be restored using the Gating BPM button.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5327438 IVY 3150B ECG Monitor (equivalent to 5235192 IVY Monitor 3150A which replaces 5112819 IVY 3150 ECG EKG Monitor) See LD for details IVY 3150B ECG Monitor, replaces 3150A and 3150 older monitors. Part #: 5327438 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IVY Standalone Monitor for IVY Cardiac Trigger Monitor 3150B. Includes operators manual. OEM Part number 3150-B FRU: Yes Status: Orderable Comments: Per ECO 2063000, 212008898, 1/8/09. Part #: 5235192 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IVY Standalone Monitor for IVY Cardiac Trigger Monitor 3150A OEM Part Number 3150-A FRU: Yes Status: Orderable Part #: 5112819 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 3150 IVY ECG MONITOR FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues