GE Healthcare · CTLightSpeed23000006450003164000245180
LightSpeed 4.X - GOC2
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50003/5/1 Scan Seq Id: 164000 TAXI data violation detected by forward error correction. Interrupt bit: 4 Violation count: 1 Scan aborted. Error: 245180 System aborted Xray-on Static scanning. Wed May 12 13:54:02 2004 Error: 245180 Host: ctpro100 Process: scanmgr File: scan_complete.cxx Line: 313 Function: System Control : Scan Control Scan Type: Xray-on Static Scan: 0/0/0 Scan Seq Id: 1000 System aborted Xray-on Static scanning. Views expected 992 Views collected 0 an updated version of this error Wed Feb 9 14:53:22 2011 Error: 245180 Host: ct04 Process: scanmgr Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 System aborted Helical scanning. Views expected 8762 Views collected 5555
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2333611 4.X-5.X 850 MB RF RECEIVER See Long Description Part #: 2333611 Part Type: GP CONSUMABLES Part Weight: 0.14 LBR Repairable: No Returnable: No Hazardous Material: N Description: 850 MB RECEIVER FRU: Yes Status: Orderable Comments: FLS - 850MB RECEIVER ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.