Pular para o conteúdo principal
GE Healthcare · CTLightSpeed178568

LightSpeed 4.X - Octane

Symptom

Scanning HW reset fails to reset OBCR

Error message

Error: 178568 Failed to establish socket connection to OBCR in 10 seconds. Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. ---------------------------------------------------------------------------- Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2260421 replaces 2260421-2 VME console power supply. Used on CTi - 4.x H3 GOC1 and GOC2 Operators Console. See Long Description for Details. 2260421 replaces 2260421-2 Part #: 2260421 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VME CONSOLE POWER SUPPLY ASSEMBLY FRU: Yes Status: Orderable Part number 2260421-2 is a comsumable assembly at EOL with 45 per year comsumsion. Part number 2260421 is the same base assembly with an added bracket and is a repairable. Do to the EOL issue and low comsumption of 45 per year we will eliminate p/n 2260421-2 and replace it with p/n 2260421 to have one repairable part. This will eliminate the need to stock two parts and create a new repair process.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues