Infinia Hawkeye 4 is unable to perform X-ray acquisition with gantry rotation.

Symptom

Infinia Hawkeye 4 is unable to perform X-ray acquisition with gantry rotation. This problem occurred during a new Infinia Hawkeye 4 installation. The gantry would rotate a few degrees, then stop. A few seconds later, an error message was displayed on the acquisition PC. This happened when trying to acquire X-ray daily QC, clinical scans, and when trying to run X-ray diagnostics that require gantry rotation.

Error message

EVENT_CHUNK_TIMEOUT The following error is logged in the CTServer error log: ERROR=Error in CTServer :\tacq\su\src\cpp\ct\hawkeyems\device\hkmsprojectionprovider.cpp(1492): EVENT_CHUNK_TIMEOUT timeout, 15000

Applies to

Models
Infinia
Product
Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
Component
5317933 (5318465 SEM) - Infinia Hawkeye 4 SEM board 5317933 (5318465 SEM) - Infinia Hawkeye 4 SEM board March 2010 New SEM P/N 5317933 Infinia II/HE4, Compatible in all systems, EXCEPT Infinia 1 Old Part Number 5175923 Infinia Hawkeye 4 SEM board (PCA500013) *X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller)
Area
Nuclear Medicine