GE Healthcare · Nuclear MedicineInfinia15000
Infinia Hawkeye 4 is unable to perform X-ray acquisition with gantry rotation.
Symptom
Infinia Hawkeye 4 is unable to perform X-ray acquisition with gantry rotation. This problem occurred during a new Infinia Hawkeye 4 installation. The gantry would rotate a few degrees, then stop. A few seconds later, an error message was displayed on the acquisition PC. This happened when trying to acquire X-ray daily QC, clinical scans, and when trying to run X-ray diagnostics that require gantry rotation.
Error message
EVENT_CHUNK_TIMEOUT The following error is logged in the CTServer error log: ERROR=Error in CTServer :\tacq\su\src\cpp\ct\hawkeyems\device\hkmsprojectionprovider.cpp(1492): EVENT_CHUNK_TIMEOUT timeout, 15000
Applies to
- Models
- Infinia
- Product
- Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
- Component
- 5317933 (5318465 SEM) - Infinia Hawkeye 4 SEM board 5317933 (5318465 SEM) - Infinia Hawkeye 4 SEM board March 2010 New SEM P/N 5317933 Infinia II/HE4, Compatible in all systems, EXCEPT Infinia 1 Old Part Number 5175923 Infinia Hawkeye 4 SEM board (PCA500013) *X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller)
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.