Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200280012

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Mouse changes state from arrow icon to hour glass icon in viewport. See Long Description for details Mouse changes state from arrow icon to hour glass icon in viewport. Intermittently the mouse changed states while in the viewports, causing havoc in trying to execute mouse clicks.

Error message

Error: 200280012, Attempt to reenter real time mouse routine See Long Description for details. Error: 200280012, Attempt to reenter real time mouse routine Wed May 30 00:58:05 2007 Error: 200280012 Host: stmarysctls Process: afViewer File: ctBoard.cxx Line: 192 Attempt to reenter real time mouse routine

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2368697-2 LINUX 3 BUTTON PS2 MOUSE See LD for details Part #: 2368697-2 Part Type: GP CONSUMABLES Part Weight: 0.22 LBR Repairable: No Returnable: No Hazardous Material: N Description: LINUX 3 BUTTON PS2 MOUSE FRU: Yes Status: Orderable Comments: FLS - LINUX DDR WORKSTATION 3 BUTTON PS2 MOUSE . THIS P/N REPLACES P/N 2330676-2 FOR THE GOC2-LINUX CONSOLE, PER PCN 220795. MS20544 ON 2/19/03. DO NOT!!!! CHANGE TO 2368697 THIS IS THE WHOLE COMPUTER! IN CASE YOU MAY NEED IT, YOU CAN PROCURE INSTEAD 2348405-2. (HP P/N Y1755A). THIS HAS BEENIDENTIFIED TO BE THE SAME AS P/N 2368697-2 (HP P/N A28314).(THE DIFFERENT HP P/NS ARE DUETO INTERNAL SUPPLY CHAIN INTEGRATION BETWEEN HP AND COMPAQ).MS20544 ON 3/11/03. PART PLACED ON CONSUMABLE COLLECTIO PER REQUEST OF CHAD VANDE HEI MS13452 030519 ALSO USED ON EXCITE II FROM MR, PER RSPL. MS20544 6/17/03.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues