LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state Error: 244634 Subsystem State; DCB:down, CCB:down, TCB:down, JEDI:down, TGP/ORP:down, All Down See Long Description for complete error Thu Mar 18 21:29:36 2010 Error: 244634 Host: sim1 Process: fwmgr File: fwmgr utils.c Line: 664 Running: System Reset Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:down TCB:down JEDI:down TGP/ORP:down
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2363101 GDAS Fuse box assembly and the 2363101-3 is the GDAS EMC Edition 2 Fuse Box ASM See long description for details. Part #: 2363101 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE BOX ASM, H4 FRU: Yes Status: Orderable Comments: FLS - FUSE BOX ASM, GDAS USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 040316 RISK: SITE DOWN FREQUENCY: LOW THIS BOX ASM CONTAINS THESE LOWER LEVEL FRU: 46-170042P22 30A, 600V FUSE 46-136343P15 SPST, ON-OFF 20A TOGGLE 2106993-5 20A 500 VOLTS .3 PDR 8537.10.9070 IM05 AWAITING MM UPDATE Part #: 2363101-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Fuse Box ASM, GDAS EMC Edition 2 FRU: Yes Status: Orderable Comments: Part added by RT GDAS RSPL, this part is also replacement for 2363101-2. 2363101-2, is NOT forward compatible and can not be used to replace 2363101-3. Please use up existing stock of 2363101-2 before crossing -2 orders to -3, ms13452 050211
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required