GE Healthcare · CTLightSpeedVCT200600033

LightSpeed 5.X (80/100kW)

Symptom

Retro reconstructions fail. Retro reconstructions fail WITH Error:200600033 Attempting to produce retro series with 3032 images is not allowed The image number in Browser or in the image queue is already 2803. posted in error log, this is most likley to occur when customer is doing retro reconstructions of cardiac gated images

Error message

Error: 200600033 Attempting to produce retro series with XXXX images is not allowed See long description for variations Error: 200600033 Host: ct01 Process: retro Recon File: UIRxToRetroRequest.cxx Line: 1627 Attempting to produce retro series with XXXX images is not allowed The image number in Browser or in the image queue is already XXXX XXXX number of images loaded in the retrorecon series that exceeded permissible SW limit

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution