GE Healthcare · CT230006527260005009
Lightspeed 1.X - Octane
Symptom
Recon self test failure on boot up. Recon self test failure on boot up
Error message
Error: 230006527 Recon Subsystem Reset FAILED See long description for full error Sun Mar 11 10:36:12 2007 Error: 230006527 Host: cbct_sbc Process: reconmgr File: rm_recon_events.c Line: 786 Function: Image Generation : SBC Processing Job Type: Not defined Scan: 0/0/0 Image: 0/0/0 Recon Seq Id: 0 Recon Subsystem Reset FAILED Error: 260005009 Timeout waiting for ICE CPU applications to come up See long description for full error Sun Mar 11 10:36:12 2007 Error: 260005009 Host: cbct_sbc Process: procICE File: iceTimeout.c Line: 103 Timeout waiting for ICE CPU applications to come up : status=0. Recon will not be available without a successful reset
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2191454-3 SCSI Cable used on LightSpeed H1& H2 Systems See LD for details Part #: 2191454-3 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCSI CABLE, DRV BOX TO S/BLKHD, HD68 FRU: Yes Status: Orderable Comments: FLS - SCSI CABLE, DIV BOX TO S/ BH HD68. REPLACES 2191454-2 PER PCN 199904. USE UP OLD FIRST.USED ON H1 AND H2. MS26561 DSM 11/7/00 *PER CT MFG - THEY NO LONGER STOCK THIS AT CT AS IT IS OBSOLETE TO CT - 1/29/03 - MS22763-JP* MUST OBTAIN FROM NORTECH LightSpeed QXi with no O2 or original H2 Systems with HSD in Dog cage: Bulkhead J51 to Drive Box (Molded) is 2191454-3
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required