GE Healthcare · CTLightSpeed290009023290009016290010005290009013290012004290009012
LightSpeed 5.X (60kW)
Symptom
Image recon is failing intermittently - document in RFS - critical issue See long description for details. Image recon is failing intermittently.
Error message
Error: 290009023 Resetting IG (1,2,3) because it has reached max consecutive retries: 4 See Long Description for complete error. Error: 290009023 Resetting IG (1,2,3) because it has reached max consecutive retries: 4 Sun May 14 13:40:24 2006 Error: 290009023 Host: darc Process: recon_control File: IG_interface_list.cxx Line: 1272 Resetting ig2 because it has reached max consecutive retries: 4 Note: This could be IG1, IG2, IG3 Error: 290009016 Recon is going to SHUTDOWN mode due to XXXXXX. See Long Description for Error variation Tue Jul 12 14:31:49 2005 290009016 4 darc recon_control recon_control_cb.cxx 464 Recon is going to SHUTDOWN mode due to no IG can be used to perform reconstruction. Mon Aug 31 23:13:32 2009 Error: 290009016 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 6029 Recon is going to SHUTDOWN mode due to maximum number of consecutive suspends reached. Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminated because IG dies Fri Nov 16 19:22:57 2007 Error: 290010005
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2362872 Westville IG Module used on GOC3 and GOC4 consoles. See Long Description for details. 2362872 Westville IG Module used on GOC3 and GOC4 consoles. Part #: 2362872 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 19.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IMAGE GENERATION COMPUTER - GRE FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell IG 5147443 to replace a Westville IG (2362872), you MUST order p/n 5193989 software release 06MW29.7, and perform a LFC. See Service note SN06101010A. PET must receive the Westville IG.(12/12/06 212001261) *** 2ND OPINION NOTES *** See the following URL for additional command line diagnostic troubleshooting: http://gemsvst4p.med.ge.com:7005/facets/CT/GRE%20command%20line%20diag%20test.pdf?
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.