GE Healthcare · CTHiSpeed110021105711003

Cannot send a specific exam to PACS

Symptom

Cannot send a specific exam to PACS Cannot send exam to PACS with Error: 11057 Translation Failure. Genesis . Grp: 32 , Elm:82

Error message

Error: 11002 DCM low level error: Application error (505): Application error (172) Tue Jul 11 15:47:11 2006 Error: 11002 Host: cass Process: dcserver File: dcscommon.c Line: 228 DCM low level error: Application error (505): Application error (172)

Error: 11057 Translation Failure. Genesis . Grp: 32 , Elm:82 See LD for full error Error: 11057 Translation Failure. Genesis . Grp: 32 , Elm:82 Wed Feb 15 18:54:24 2006 Error: 11057 Host: ctge01 Process: dcserver File: dcspush.c Line: 784 Translation Failure. Genesis . Grp: 32 , Elm:82

Error: 11003 Failure to get DCM event: lost transport connection (114) Fri Jul 23 08:28:45 2004 Error: 11003 Host: lightspeed04 Process: dcserver File: dcspush.c Line: 2514 Failure to get D From viewlog: Fri Jul 23 08:28:45 2004 Error: 11003 Host: lightspeed04 Process: dcserver File: dcspush.c Line: 2514 Failure to get DCM event: lost transport connection (114)

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1
Area
CT

Full cause & step-by-step resolution