Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 260101008 The ORP failed to connect with the CCB sub-system in the allotted time. Check to see if subsystem is alive.Timeout value: 4000ms error 260101008 The ORP failed to connect with the CCB sub-system in the allotted time. Check to see if subsystem is alive.Timeout value: 4000ms Open LD to see error example Error: 260101008 The ORP failed to connect with the CCB sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS This was followed by Error: 244540 Immediate Request cannot be honored since the firmware is in the DOWN state. Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 See long description for details 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 Part #: 5126615 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASTEC Digital PS Assembly FRU: Yes Status: Orderable Comments: Added per RT / GDAS RSPL
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.