GE Healthcare · CTLightSpeed24505036505200001821
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 245050 Error verifying that Data Acquisition is ready for scanning Open Long Description to see error examples Error: 245050 Error verifying the Data Acquisition is ready for scanning. More than one error text exists for this error below: Fri Jul 1 16:51:14 2005 Error: 245050 Host: ct02 Process: scanmgr File: start_scanning.cxx Line: 231 Function: System Control : Scan Control Scan Type: Not scanning Scan: 483/1/0 Scan Seq Id: 1 Error -1 verifying that Data Acquisition is ready for scanning Wed Mar 3 16:56:16 2010 Error: 245050 Host: uhctlspro Process: scanmgr File: start_scanning.cxx Line: 250 Function: System Control : Scan Control Scan Type: Not scanning Scan: 36505/1/0 Scan Seq Id: 0 Error 200001821 verifying that Data Acquisition is ready for scanning
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2260421 replaces 2260421-2 VME console power supply. Used on CTi - 4.x H3 GOC1 and GOC2 Operators Console. See Long Description for Details. 2260421 replaces 2260421-2 Part #: 2260421 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VME CONSOLE POWER SUPPLY ASSEMBLY FRU: Yes Status: Orderable Part number 2260421-2 is a comsumable assembly at EOL with 45 per year comsumsion. Part number 2260421 is the same base assembly with an added bracket and is a repairable. Do to the EOL issue and low comsumption of 45 per year we will eliminate p/n 2260421-2 and replace it with p/n 2260421 to have one repairable part. This will eliminate the need to stock two parts and create a new repair process.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.