Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18242615403

LightSpeed 4.X - GRE

Symptom

Customer was not able to remote tilt gantry. In many cases , gantry will tilt from the Gantry Controls.

Error message

Error: 182426: Gantry Touch Sensor FAULT. Console Pushbutton Remote Tilt is DISABLED See Long Description for Full Error Tue Jan 10 17:42:00 2006 Error: 182426 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 97 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7855/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:41:59:940 Log Series: 197 The Gantry Touch Sensor FAULT. Console Pushbutton Remote Tilt is DISABLED. (Gantry Sensors - FORWARD Touch Sensor Error(s)). Gantry Controls are running in FAULT MODE. Current Position 15403 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down). (Elev Cmd: 1 TouchSensor: 0x5000).

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required