Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600073915010232719

LightSpeed 5.X (80/100kW)

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error # for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply Wed Feb 22 08:50:01 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 50102/3/0 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply . Measured: 1590 mV, Expected range: 4750 to 5250 mV Tue Aug 22 15:51:09 2006 260007391 4 orp orp_subsys pwrtst.c 90 Tue Aug 22 15:51:08 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 32719/3/0 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply . Measured: -4748 mV, Expected range: -4751 to -5249 mV

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5133765 GDAS Power Supply Upgrade Kit: to Astec Power Supplies See the LD for FEMC part comments 5133765 GDAS Power Supply Upgrade Kit Part #: 5133765 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS Power Supply Upgrade Kit: to Astec Power Supplies FRU: Yes Status: Orderable This part is Service replacement Kit for GDAS power supplies 2330352-4 and 2330352-3. Single kit will replace BOTH analog and digital power supplies at same time. ms13452 050420 Note: (1/2/06 SMF) Comments from Dennis L.Badgerow, The Kit is shipped in a box the size of an x-ray tube, only weighs 110 lbs but is really big should ship directly to the site if possible. Next, the directions say one FE for two hours, it took two FE's nearly five hours with no interruptions.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.