GE Healthcare · CTHiSpeed18394965001
Scanner Hardware Stopped Scan message during axial scan
Symptom
Scanner Hardware Stopped Scan message during axial scan. Long Description for Feedback Only. Scanner Hardware Stopped Scan message during axial scan
Error message
Error: 183949 Could NOT CLOSE Gantry (Axial) Loop Contactor See LD for full error CT01 Host: STC Proc: Axial Error: 183949 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor Another example: Tue May 24 08:41:21 2005 Suite: CT02 Host: STC Proc Axial Error: 183949 File: ax_error.c Method No Method 1114 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 65001/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:21:940 Log Series: 60 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e28. EN 401 SR 442
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-186852P2 A1K2 contactor See long description for details and picture Part #: 46-186852P2 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: RELAY (K2 BACK-UP CONN.) FRU: Yes Status: Orderable Comments: FLS - ALSO AVAILABLE AS 46-186852P2UUSE UP THAT STOCK FIRST! SWITCH ON SIDE OF RELAY= 46-296690P1 A4K2 PF8 TWICE FOR FE ASSIST COMMENTS!!! FOR FE ASSIST PF8 AGAIN!!! FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PARTS WHEN AVAILABLE., IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. FLS - FE ASSIST MS6588 10/2000 46-186852P2 = AXIAL RELAY ARE YOU ORDERING THE CORRECT PART?? FOR CT PDU THE AXIAL RELAY ASMIS MADE UP OF TWO RELAYS, THE 46-186852P2 & 46-296690P1 P1 RELAY MOUNTED ATOP OF P2. 46-186852P2 = AXIAL LOOP CONTACTOR RELAY. 46-296690P1= AUXILIARY CONTAC T RELAY. SEE CTI FRU MANUAL PAGE 5-5 REVISION 26 AND ABOVE, OLDER REV DO NOT SHOW THE AUX RELAY. DO YOU STILL WHAT TO ORDER THIS PART??
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.