Symptom

Scanner keeps freezing up and would not go past scout acquisition Scanner keeps freezing up and would not go past scout acquisition.The below warning has also been seen in the operator message bar: "Not Enough Image Space Please Remove Images" Scanner keeps freezing up and would not go past scout acquisition The below warning has also been seen in the operator message bar: "Not Enough Image Space Please Remove Images"

Error message

Error: 200520015 Updating image space, received a bad status code: See the long description for different status codes Sat Jan 22 18:23:06 2005 Error: 200520015 Host: chr09bc1cti Process: environmentMgr File: ImageSpaceFormMgr.c Line: 117 Updating image space, received a bad status code: 200000416 Reason: Response from SdC timeout you may also see this error with different status codes, such as Fri Jul 28 05:12:48 2006 Error: 200520015 Host: vct1 Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: -1 Reason: OS error : -1 --- Unknown error 4294967295 Wed Nov 16 08:23:54 2005 Error: 200520015 Host: gect_qxi Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: 200000415 Reason: Invalid data from SdC.

Error: 210000407 Image space query failed. See Long Description for more details Mon Feb 27 07:19:00 2006 Error: 210000407 Host: csbipct2 Process: dms File: Initialize.cxx Line: 1085 Image space query failed. Error code: -1

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.