GE Healthcare · CT
Sytec Synergy Series
Symptom
Unable to scan with XG errors
Error message
ER-117 XG Fail - contains information on all XG ER117 errors See Long Description ER-117 XG Fail ER–117 ER – 117 : XG Error 00000XYY Error Code X = _ 0: Test Mode 1: Stand–by 2: Preparation Acting 3: Generator Ready 5: Exposure 7: Transition Step to Stand–by ER–117 (continued) ER – 117 : XG Error 00000XYY Error Code YY = _ X01: No DC550V X02: 15V_Safe Error X03: Over Voltage Capacity X11: Rotor Controller Error X12: Rotor Acceleration Error X13: Rotor Feedback Test X14: Rotor Fault Detect Test X16: Low Current Error X18: Unbalance Error X21: Over Filament Current 021: Feedback Test X22: No Filament Current 022: Fault Detect Test X23: Bad Relay Return 023: Small Focus Test X24: No Filament Voltage 024: Large Focus Test X25: Filament Control Error 526: MA measure Error 031: Feedback Test 032: Fault Detector Test 033: Close Loop Test 531: MA Accuracy Error 532: KV Accuracy Error 533: Spit Error 534: Exposure Backup Timer Error 535: Anode Over mA 536: Cathode Over mA 537: Anode Over kV 538: Cathode Over kV 539: Anode Shoot Through 540: Cathode Shoot Through 541: Anode Over Current 542: Cathode Over Current X43: HV ON with Error 544: KV Unbalanced Error X48: Generator Safety Loop Open X
Applies to
- Product
- Sytec Synergy Series - Synergy S, Synergy, Synergy Plus, Sytec SRi/Legato XMj CT Sytec Synergy Series - Synergy S, Synergy, Synergy Plus, Sytec SRi/Legato XMj CT Sytec Synergy Series - Synergy S, Synergy, Synergy Plus, Sytec SRi, Legato XMj For Synergy S to Synergy Upgrade Kit (2185628) OC Label (2108440) GANTRY Label (2108443). For Synergy to Synergy Plus Upgrade Kit (2185630) GEMS–E: OC Label (2108440) GANTRY Label (2108443) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3121T / D3122T Xray tube used on Synergy and Sytec SRI systems (MX135) See long description for more information. D3122T Xray tube used on Synergy systems (MX135) Part #: D3122T Part Type: GP TUBES Part Weight: 187.39 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE SYTEC SRI AND SYNERGY FRU: Yes Status: Orderable Comments: FLS - D3122T = SYTEC SRI TUBE, MX135SYNERGY ++++++++++++++++ IMPORTANT NOTE FOR CARES-CPD++++++++++++ALSO FOR CT PACE MAXIRAY EQUIVALENT IS D3129T DIMS (LWH) 37 X 27 X 28 AM SOURCE = MEXICO PF8 FOR ADDITIONAL FE ASSIST COMMENTS!!!!!!!!!! FLS - FE ASSIST 981108 MS6588 D3122T = SYTEC SRI TUBE, MX135 MAXIRAY EQUIVALENT IS D3129T ** PLEASE CHECK FOR ANY SIGNS OF LEAKAGE FROM THE HIGH VOLTAGE TANK PRIOR TO TUBE REPLACEMENT. PRIOR TO REPLACEMENT OF TUBE BE ADVISED THE HV TANK HAS A HIGH FAILURE RATE REFER TO YMS-SN052 HV TANK = 2116229
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.