Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2600066055000026000680350079

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260006605 Could not send axial drive message in the allotted time. Axial serial link is down. See the LD for the full error code Error: 260006605 Could not send axial drive message in the allotted time. Tue Nov 16 14:20:36 2004 Error: 260006605 Host: STC Process: Axial File: scanbusDriver.c 1.5 Line: 523 Function: AXIAL CONTROL : axial communications Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:20:37:325 Log Series: 70 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices, ..

Error: 260006803 Axial Motor Control (AMC) logic could NOT access the AXIAL drive via the CAN network. See long description for complete error Error: 260006803 Host: STC Process: Axial File: amc_exec.c Line: 3077 Function: AXIAL CONTROL : axial communications Scan: 50079/4/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 05:23:06:435 Log Series: 505 Axial Motor Control (AMC) logic could NOT access the AXIAL drive via the CAN network.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.