Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260006817 Axial Motor Control (AMC) Logic detected Gantry (Axial) MOTION without Command Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1 See the long description for the full error code Mon Dec 6 09:27:53 2004 Error: 260006817 Host: STC Process: Axial File: amc_mon.c Line: 987 Function: AXIAL CONTROL : axial control Scan: 23391/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:27:53:795 Log Series: 829 Axial Motor Control (AMC) Logic detected Gantry (Axial) MOTION without Command Current Position: 108.7 degs Commanded Position: 90.0 degs Degrees Tol: +/- 1 Retry Rx to verify probable Gantry Balance or Brake or Encoder Hardware. problem.

Error: 260006815 Axial Motor Control (AMC) Logic could did NOT position the Gantry within tolerance. See long description for full error Mon Dec 6 09:27:53 2004 Error: 260006815 Host: STC Process: Axial File: amc_exec.c Line: 1494 Function: AXIAL CONTROL : axial control Scan: 23391/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:27:54:170 Log Series: 830 Axial Motor Control (AMC) Logic could did NOT position the Gantry within tolerance. Current Position: 108.7 degs Commanded Position: 90

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.