Pular para o conteúdo principal
GE Healthcare · CTHiSpeed245502

HiSpeed CT/i 4.x with SGI Indigo Console, HiSpeed Advantage Gantry

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 245502 Failure in executing high speed disk pre-scan write test See Long Description for complete error Error: 245502 Failure in executing high speed disk pre-scan write test. Thu Mar 27 21:31:21 2008 Error: 245502 Suite: CT00 Host: SBC0 Process: dataacq File: DataAcqEventHandler.m 1.35 Method: verifyAcqHw Line: 1151 Function: Data Acquisition : SBC Processing Scan: 0/0/0 Type: Not scanning Scan Seq: 0 Failure in executing high speed disk pre-scan write test.

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2226715-3 CTI 18GB ST318418N (replaces 2226715-2 CTI 9 GB) Scan Data Disk, Used on Indigo OC, sbc local scan disk and sbc scan data disk -and Octane scan data disk See LD for more information and Picture Part#: 2226715-3 replaces 2226715-2 ( also shown below for reference ) Part #: 2226715-3 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 18GB (DSTRK 9GB) FRU: Yes Status: Orderable Comments: *** OBSOLETE PART CASE 5523893 *** This part has been identified as obsolete and no short-term solution has been identified. FE must identify and order a higher or lower level spare part as required. Contact your support engineer, on-line engineer or Parts ID team should you require support. 212012173 10-2-06 11/19/04 (Shawn Fox 212001261) Please return your defective Drive's to: Babar Maqsood 3000 N. Grandview Blvd Mailstop W-1250 Waukesha, WI 53188 FE ASSIST MS6588 MARCH 2002 2226715-3 = ST318418N 18GB DRIVE. 2226715-3 IS A DESTROKED DRIVE 2226715-3 CAN BE USED IN THE FOLLOWING LOCATION: CTI /W INDIGO -3 WORKS IN OC SBC/LSD & SCAN DATA DISK. CTI /W OCTANE -3 WORKS IN SBC/LSD & SCAN DATA DISK. WHEN USING A -3 AS A CTI INDIGO OC DISK PLEASE READ: FOR CTI INDIGO OC DISK A SOFTWARE SCRIPT IS NEEDED TO LOAD SOFTWARE. THE SOFTWARE SCRIPT CAN BE FOUND ON SUPPORTCENTRAL OR IN THE CLARIFY PROBLEM SOLUTION DATABASE. 212002185 05/10/2007 22:28 GMT CTI 9 GB Scan Data Disk, part# 2226715-2 Used on Indigo OC, sbc local scan disk and sbc scan data disk - and Octane scan data disk picture below Part #: 2226715-2 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCAN DISK SBC SYSTEM 9.2 GB USE WITH 2204245-3 CD ROM 6.2 OS FRU: Yes Status: Orderable Comments: 11/19/04 (Shawn Fox 212001261) 2226715-2 = NEW 9 GB DRIVE SEAGATE ST39216N. 2226715-2 IS USED ON INDIGO (CTI4.4 & FORWARD) SBC LOCAL SCSI DISK AND INDIGO OC DISK. 2226715-2 IS USED ON OCTANE SYSTEMS (CTI 6.1 & FORWARD) INTHE SBC LOCAL SCSI DISK. Below is a pic of the disk location from CT Hispeed Systems FRU Manual, Direction 46-018316 Rev 37 Octane OC below
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.