Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed1762902584

BrightSpeed Select (Hivalue Lite) 4/8/16 Slice Scanner

Symptom

MOD Save or Restore functions fail. MOD Save or Restore functions fail Troubleshooting There is no media in the drive message appears in a attention box during a MOD Save or Restore of with Error: 10191 Failed save exam. posted in the Message Log. The fllowing pop-up message appears on screen. Problem is consistent If a different disk is used then the error occurs again when trying to save, restore, or label

Error message

arc_attachDMOD:Checksum failure in arslog in directory /usr/g/ctuser/logfiles Recovery Failed status 255 for MID:1762902584::2003:01:09:21:26:19GMT:CST Fri Jan 10 17:02:16 2003 Recovery: Recovery took 38.4743 seconds cd /usr/g/ctuser/logfiles arslog arc_attachDMOD:Checksum Sometimes the recovery fails after 40 seconds but upon reattaching the MOD the no corrupt media is detected. "Recovery Failed status 255 for MID:1762902584::2003:01:09:21:26:19GMT:CST Fri Jan 10 17:02:16 2003 Recovery: Recovery took 38.4743 seconds OCT 15 DICOM issue... MESSAGE from Process 4061>> Tue Oct 15 11:25:37 2002 (ArchiveDevice.c:3303) Attach in progress Tue Oct 15 11:25:40 2002 RetrieveDicomDirRec: grp: 20 elm: 32 DS or AN string is formatted incorrectly Tue Oct 15 11:25:40 2002 arc_attachDMOD:Checksum failure Tue Oct 15 11: arslog: END MESSAGE from Process 5997>> Fri Jan 10 16:06:31 2003 (arsjob.c:341) [Arserver]> init of arsjob connected to NETserver MESSAGE from Process 5997>> Fri Jan 10 16:06:31 2003 (ArchiveDevice.c:4615) [Arserver]> Resetting ARS SONYSMO-F551 - $Id: @(#)ArchiveDevice.oN32 vobadm Thu Jun 13 03:16:41 CDT 2002 /view/IRIX646.5__build__R2.9j $ MESSAGE from Process 5997>> Fri Jan 10 17:

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5167410 SCSI Tower with MOD and DVD drives for GOC series consoles (replaces 2368838 ) See Long Description for details Part #: 5167410 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Peripheral Tower, dual-bay, MOD-DVDram FRU: Yes Status: Orderable Comments: ** FE ASSIST ** If ordering 5167410 (new) SCSI tower to replace 2368838 (old) SCSI tower for issues not related to the MOD, contact FMI Admin (Sahu, Shankar, sso 501375480 ) to order FMI 25408 (kit# 5317459) instead. FMI 25408 re-uses MOD, so if that is the failure, then continue with 5167410 order.(212001261 11/10/08) *** Second Opinion *** Reconfig should be tried prior to declaring DOA/FOA part (212001261 7/10/09) Per ECR 2030398, 8/2/06, MS11892.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.