LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scan hardware stopped scan with DAS Board errors. Scan hardware stopped scan DAS Board errors Scanner Hardware Stopped Scan message appears in the operator console area during Axial Scan data acquisition with Error: 260001023 The DCB has notified the system of an error condition which prevents scanning and error: 260007402-Converter Board(s) Asserted a Board FAULT cards 76,78,80,82,84,86,88,90,92,94,96 and Error #260007774 list of converter cards from previous error message Problem was intermittent Then after time the errors would occur more often and eventually the problem was consistent [insite@erct ~]$ /usr/g/insite/bin/viewlog Sun Aug 29 Error: 260001023 Host: OBC Process: HOP File: hop erlog.c Line: 367 Function: No System Function Reported Scan: 50000/1/2 Type: Axial Exception Level: Pri/Most Severe Time: 08:37:10:145 Log Series: 120 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002358 Notification level: 5 Sequence number: 11
Error message
Error: 260132793 Digital Modules Have Not Been Detected By The Data Acquisition System See Long Description for complete error. Mon Feb 6 09:42:29 2006 Error: 260132793 Host: orp Process: orp_subsys File: IfbManager.cxx Line: 318 Mon Feb 6 09:42:28 2006 Host : Das Ermes # : 260132793 Exception Class : Abort Severity : Primary File : IfbManager.cxx Line# : 318 Function : No System Function Reported Scan Type : Static Scan: 50089/1/1 The Following Digital Modules Have Not Been Detected By The Data Acquisition System. See Previous Messages From Last Reset For Detailed Fault Information. Digital Modules: 37-40 (or some equivalent number)
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5196402 FRDM for Outer Modules used on VCT HDAS-64 See Long Description for details 5196402 FRDM Module for OUTER SLOTS used on VCT HDAS-64 Part #: 5196402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: FIELD REPLACEABLE MODULE KIT - OUTER SLOTS FOR HALO CONVERTER CARD FRU: Yes Status: Orderable
- Area
- CT