GE Healthcare · CTLightSpeed2600073915010232719

LightSpeed 4.X - GOC2

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error # for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply Wed Feb 22 08:50:01 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 50102/3/0 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply . Measured: 1590 mV, Expected range: 4750 to 5250 mV Tue Aug 22 15:51:09 2006 260007391 4 orp orp_subsys pwrtst.c 90 Tue Aug 22 15:51:08 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 32719/3/0 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply . Measured: -4748 mV, Expected range: -4751 to -5249 mV

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2291847 Power Supply D Case W/ AC&DC Harness. See long description for details Part #: 2291847 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 3.01 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PWR SPLY ASM D-CASE W/HARNESS H3 FRU: Yes Status: Orderable Comments: FLS - PWR SUPPLY D-CASE W/AC&DC HARNESS. USED ON WARP3. SUPPLIER IS CT MAN Item # 2 below
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.