GE Healthcare · CTHiSpeed368683288836961369603314436973
After replacing O2 system module, get HINV boot up error, different hardware inventory vs the default file, FLASH PROM version difference
Symptom
After replacing O2 system module, get HINV boot up error, different hardware inventory vs the default file, FLASH PROM version difference After replacing O2 system module, get HINV boot up error, different hardware inventory vs the default file, FLASH PROM version difference.
Error message
usr/g/service/log/poweron.log file has an hinv error: Caution : we get diffarent hardware inventory vs. the default file. FLASH PROM version 4.1x See Long Description for a complete printout of the failing poweron.log file usr/g/service/log/poweron.log file has an hinv error: Caution : we get different hardware inventory vs. the default file. FLASH PROM version 4.1x IRIS% more poweron.log hinv test : Error *** PCI Device List *** - Adaptec SCSI : PCI Adapter ID (vendor 36868, device 32888) pci slot 1 - Adaptec SCSI : PCI Adapter ID (vendor 36868, device 32888) pci slot 2 - NPR : PCI Adapter ID (vendor 4277, device 36961) pci slot 5 - DBPCI : PCI Adapter ID (vendor 4277, device 36960) pci slot 6 - Adaptec SCSI : PCI Adapter ID (vendor 36868, device 33144) pci slot 4 - Serial Exp : PCI Adapter ID (vendor 4277, device 36973) pci slot 7 *** Caution : we get different hardware inventory vs. the default file. *** Please check the hardware enviroment. *** Diffarence(s) form the default file is(are) : *** 8c8 < FLASH PROM version 4.17 --- > FLASH PROM version 4.14 child process exited abnormally System Disk : Ok sc0d1l0: dsopen(sc0d1l0,0) dsopen(/dev/scsi/sc0d1l0,0) =>cp 1001a018, dsp 1001
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Area
- CT
Full cause & step-by-step resolution