RXi Cannot Fluoro in apps, no error reported in applications.
Precision RXi Cannot Fluoro in apps, no error reported in applications due to a defective cable between the CPI Generator and the Infimed Digital.
Symptom
RXi Cannot Fluoro in apps, no error reported in applications. See Long Description for details. Cannot Fluoro in apps, no error reported in applications. Can Fluoro in Digital service mode, but get poor image and techniques very low. Iris cal not successful. ABS signal is a constant 5 Volts on test points TP8-9 on GIB board (typical is approx 2 Volts during Fluoro, depending on Fluoro curve selected). Digital spots are fine in both apps and service modes.
Applies to
- Models
- Precision
- Product
- Precision RXi
- Component
- Infimed Digital CPI Generator
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug
- Precision 500D Aurora, RCIM E-stop circuit is difficult to troubleshoot because there is no overall interconnect schematic from the RCIM operator console to the Corona 2 board in the System Cabinet.
- Precision 500D Aurora system with Magic PC.
- P500D Siemens Collimator lamp voltage adjustment instructions missing due to documentation issues.