GE Healthcare · CTBrightSpeed260115209515825516693260115196516886
BrightSpeed 4/8/16 slice
Symptom
Table will not move up, especially when in full down position. See long description Table will not raise or move up, especially when it's in the full down position and this could be intermittent along with noises noticed by the customer.
Error message
Error: 260115209 Elevation stopped at invalid position. see long description for details Sat Mar 15 06:13:29 2008 Error: 260115209 Host: tgp Process: tgp subsys File: mtcb emc motion check.c Line: 294 Sat Mar 15 06:13:30 2008 Host : Table Sub-System Ermes # : 260115209 Exception Class : Secondary Severity : Pri/Soft File : mtcb emc motion check.c Line# : 294 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation stopped at invalid position. expected target position = 515825 um actual stopped position = 516693 um Error: 260115196 Elevation lower limit switch is activated. see long description for details Sat Mar 15 06:13:29 2008 Error: 260115196 Host: tgp Process: tgp subsys File: mtcb emc isrfunc.c Line: 79 Sat Mar 15 06:13:29 2008 Host : Table Sub-System Ermes # : 260115196 Exception Class : Secondary Severity : Pri/Soft File : mtcb emc isrfunc.c Line# : 79 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation lower limit switch is activated. Position is 516886 um from the floor.
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.