GE Healthcare · CTLightSpeedVCT1855533486521538
CT Performix tube rule detected by PDS
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 185553 - Alarm status: 20H (The HEMRC drive has detected a Line Loss. See Long Description for the complete error Tue Aug 10 16:44:06 2004 Error: 185553 Host: OBC Process: Generator Function: XRAY GENERATION : Tube Rotor Control Scan: 34865/2/1 Type: Helical Exception Level: Pri/Most Severe The HEMRC drive has detected a Line Loss. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: X-Ray tube stator, HEM-IT, HEMRC power supply, .. Alarm status: 20H Or Fri Mar 30 15:21:20 2007 Error: 185553 Function: XRAY GENERATION : Tube Rotor Control Scan: 21538/1/2 Type: Scout Exception Level: Pri/Most Severe The HEMRC drive has detected a Line Loss. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: - X-Ray tube stator, - HEM-IT, - HEMRC power supply, ... Alarm status: 26H . Note: Alarm Status 26H instead of 20H. This error concept should have additional information for the meaning of the various Hex codes. 20 to 21: Undefined Fault This fault code is undefined and may indicate a defective Drive. Retry operation. If problems persist contact the factory for instructions. or 20:
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2115199 HEMRC - HIGH EFF MTR ROTOR CTRL used on LightSpeed 2.x (Plus), 3.x (Ultra) and 4.x (16 Slice) Systems See Long Description for more info and picture Part #: 2115199 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 7.28 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HIGH EFF MTR ROTOR CTRL FRU: Yes Status: Orderable Comments: FLS - HIGH EFF MTR ROTOR CTRL PURCH SPEC USED ON FAST RHAPSODE 7/1/97 M3 FW28 GWENN 08992 THIS IS THE AC DRIVE FOR HEMRC THIS IS OUT-SOURCED TO A 3RD PARTY FOR ALLEN-BRADLEY. NOT DO NOT DISCOUNT - PER SHAKUN 10/30/97 GWENN MS08992 6/02 BOT FROM HOLT ELECTRIC A/B P#: 1336SBRF50-AN-EN901-MX1-L4 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required