GE Healthcare · CTLightSpeedVCT260111601200109130200000018
Step and Shoot (SmartStep) hand controller not functioning
Symptom
Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.
Error message
Error: 260111601 IHHC reset timeout. See LD for complete error Thu Oct 20 12:41:32 2005 Host : Table Gantry Processor Ermes # : 260111601 Exception Class : Abort Severity : Primary File : hc_exec.c Line# : 766 Function : No System Function Reported Scan Type : None/Unknown Scan: 4628/4/0 IHHC reset timeout. Error: 200109130 Failed to send Enable IHHC PREP request to Firmware. Error: 200109130 Failed to send Enable IHHC PREP request to Firmware, status = 200000018 Mon Aug 2 11:56:15 2004 Error: 200109130 Host: ctgep01 Process: scanRx File: uifIpcMgrScanRx.c Line: 6463 Function: Data Acquisition : OC Processing Failed to send Enable IHHC PREP request to Firmware, status = 200000018, seq_num = 3053, prep_state = 1.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.