Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1828641828691828732229871

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

One Gantry Control Panel says "OFF" and buttons will not respond. See long description for details Either the Left or Right Gantry Control Panel says "OFF" and the buttons will not respond. Panel appears to be bad.

Error message

Error: 182864 GUIF CAN Node Failed to Respond to Watchdog. Node is DOWN. See long description for complete error Mon Feb 14 06:27:44 2005 Error: 182864 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 426 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:27:44:645 Log Series: 32 GUIF CAN Node Failed to Respond to Watchdog. Node is DOWN. Node Id: 2 , Node Type: UNKNOWN: Board Number and Serial Number are UNRELIABLE. Board Number: 0 Serial Number: 0.

Error: 182869 GUIF CAN Gantry Control Timed Out on Updating its Button Status to the TNC. See long description for details Mon Feb 7 06:20:25 2005 Error: 182869 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:20:25:645 Log Series: 32 GUIF CAN Gantry Control Timed Out on Updating its Button Status to the TNC. Node Id: 5. Board Number: 0 Serial Number: 0.

Error: 182873 GUIF CAN Communications Bus Error. Node Id: 0 , Node Type: TNC (ETC I/F). Thu May 5 20:12:25 2005 Error: 182873 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 517 Function: N

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5139590 Control assembly left side gantry. ( replaces part# 2256729 and 2330695 ) – used on H2 & HP gantry’s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) FEMC info in Long Description. 5139590 Control assembly left side gantry. ( replaces part# 2256729 and 2330695 ) – used on H2 gantry’s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) Part #: 5139590 Part Type: GP CONSUMABLES Part Weight: 4.41 LBR Repairable: No Returnable: No Hazardous Material: N Description: H POWER OPERATOR CONTROLS LEFT SIDE FRU: Yes Status: Orderable If part number 2330695 is ordered it will automaticaly cross to part 5139590 Old part info 2256729 Control assembly left gantry Part #: 2256729 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CONTROL ASM LEFT H2 GANTRY FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required