GE Healthcare · CTLightSpeedVCT18554850525118336
CTx00088 CT Performix tube rule detected by PDS
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 185548 The HEMRC operating frequency is below the minimum value. See Long Description for full error Error: 185548 Suite: CT01 Host: OBC Process: Diag Exec File: hemrc_manager.c 1.5 Method: gen_rotor Line: 859 Function: Diagnostics XRAY GENERATION : Tube Rotor Control Test Name: HEMRC Manual Unique ID: 50525 Iteration: 1 Exc. Level: Hard Time: 118336 Error Type: Hardware The HEMRC operating frequency is below the minimum value. Actual frequency: 219.172 Minimum frequency limit: 249.995
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-195120G21 HV CABLES 3 PIN 7 FT See LD for details Part #: 46-195120G21 Part Type: GP CONSUMABLES Part Weight: 3.19 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLES 3 PIN 7 FT FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.