CTx00088 CT Performix tube rule detected by PDS
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 183316 (The HEMRC failed to reach the Run speed in the allotted time) Open Long Description for more details Tue Aug 10 17:58:58 2004 Error: 183316 Host: OBC Process: Generator File: hemrc_manager.c 1.14 Function: XRAY GENERATION : Tube Rotor Control Scan: 34867/1/1 Type: Scout Exception Level: Pri/Most Severe The HEMRC failed to reach the Run speed in the allotted time
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-195120G21 HV CABLES 3 PIN 7 FT See LD for details Part #: 46-195120G21 Part Type: GP CONSUMABLES Part Weight: 3.19 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLES 3 PIN 7 FT FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution