GE Healthcare · CTHiSpeed183416
HiSpeed CT/i 4.X with Indigo and HSA Gantry
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 183416 Aperture did NOT find the Z Pulse position in time. See LD for more details Fri Jun 23 10:23:53 2006 Error: 183416 Suite: CT01 Host: OBC Process: Collimator File: move_aper.c 1.20 3/3/99 Method: No Method Line: 140 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/1 Type: Static Exception Level: Pri/Most Severe Time: 10:23:53:360 Log Series: 39 Aperture did NOT find the Z Pulse position in time. (No Hardware Interrupt) Time allowed = 4000 msec.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296300G4 and 46-296300G5 EMC COLLIMATOR -used on CTi and compatible to HSA See LD for more info 46-296300G4 and 46-296300G5 EMC COLLIMATOR -used on CTi and compatible to HSA picture 46-296300G4 EMC COLLIMATOR Part #: 46-296300G4 Part Type: GP REPAIRABLES Part Weight: 19.54 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 860-950117 EMC COLLIMATOR FRU: Yes Status: Orderable Comments: FLS - FE: READ SERVICE NOTE T-1379 5/96 GVR T3202GR MPS INFO: B7999KK CONTAINS G4 COLLIMATOR PLUS S/W. MFG PLANIS TO DEINSTALL THE G3, HAVE FE RETURN, UPGRADE, AND USE INADDITIONAL KITS PER R. KOPP. 10/95 FLS - FE ASSIST MS6588 5/2000 46-296300G4 = GRAPHITE FILTER COLLIMATOR. IF NO STOCK IS AVAILABLE ORDER 46-296300G5. THE G5 COLLIMATOR IS COMPATIBLE ON CTI HSARP AND HSAZ SERIES SYSTEMS. DO YOU STILL WANT TO ORDER THIS PART??? Part #: 46-296300G5 Part Type: GP REPAIRABLES Part Weight: 19.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: EMC COLLIMATOR FRU: Yes Status: Orderable Comments: FLS - COST REDUCED EMC COLLIMATOR USED ON FAST RHAPSODE 7/2/97 GWENN 08992 T3202KM 11/19/02 -BUILD OUTSOURCED FRMCT MFG TO ERDA PER GINA DUNDUNM KRAETSCH. MS12464 FLS - FE ASSIST MS6588 5/2000 T3202KM = G5 COLLIMATOR COST REDUCED EMC COLLIMATOR USED ONHSARP, HSA AND CTI SYSTEMS. THE G5 (T3202KM) WILL NOT WORKON HELIOS/LIGHTSPEED SCANNERS. DO YOU STILL WANT TO ORDER THIS PART???
- Area
- CT
Full cause & step-by-step resolution