CTx00088 CT Performix tube rule detected by PDS
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 183184 mA out of 5 % tol Open Long Description for full error Tue Jun 7 03:35:46 2005 Error: 183184 Suite: GEC2 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 924 Function: XRAY GENERATION : MA Loop Scan: 65001/11/14 Type: Axial Exception Level: Sec/Soft Time: 03:35:45:600 Log Series: 223 mA out of 5 % or 2 mA tol Selected technic: 140 kV, 56 mA Tot. samples = 4 Tot. mA samples above tol. = 0, below tol. = 1 Avg. of mA samples above tol. = 0.0 mA, below tol. = 0.2 mA Avg. of tot. mA = 26.7 mA
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2143147 KV BOARD T3202HK 2143147 KV BOARD T3202HK 2143147 KV BOARD Part #: 2143147 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: KV BOARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution