GE Healthcare · CTHiSpeed183139
Scan Aborts and will not resume
Symptom
Scan Aborts and will not resume
Error message
Error: 183139 Rotor controller fault detected: Capacitor overvoltage. HVDC Bus capacitor voltage > 375. See LD for complete error Sun Sep 11 01:08:28 2005 183139 PNSL OBC Generator gen_abort.c 1.32 No Method 506 Function: XRAY GENERATION : Tube Rotor Control Scan: 0/0/25 Type: Static Exception Level: Pri/Most Severe Time: 01:08:28:115 Log Series: 158 Rotor controller fault detected: Capacitor overvoltage. HVDC Bus capacitor voltage > 375. Fault latch address = ffb82b hex, Bit = D6
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 45435961 CTVRC // (aka 2122768-2 see LD for why) - used on HSA, CTi and YMS Prospeed see long description Part #: 45435961 Part Type: GP REPAIRABLES Part Weight: 41.89 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CTVRC// 2122768-2 FRU: Yes Status: Orderable FLS - H&F:THIS IS THE NON-EMC CTVRC.THE MODULE HAS 8 JUMPERS WHICHWILL ALLOW CONVERSION FROM A HSA CTVRC TO A ZJ CTVRC. THE ITEM IS STOCKED WITH JUMPERS IN THE CORRECT LOCATION FOR HSA. PER HARRY LEWIS 6/94. ============================================================HQ: IF NO STOCK, YOU CAN SHIP T3202GH EMC CTVRC. T3202GH ISBACKWARDS COMPATIBLE. NEVER SHIP T3202CB IN PLACE OF T3202GH. T3202CB CANNOT BE UPGRADED. MS8365 2/95 ============================================================ HQ INFO ONLY: PCN 189116 CHANGED EQV # FOR T3202CB FROM 45435961 TO 2106148. CTVRC=CT VOLTAGE REGULATOR CONTROL.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.