Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260132919260132914

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Can't perform fastcal. Detector Rail Temperature Out Of Range

Error message

Error: 260132919 Rail Thermal Sensor Failure (OPEN) Sensor: Rail Center Channel (Left or Right also) See long description for details Wed Feb 6 22:07:35 2008 Error: 260132919 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 1728 Wed Feb 6 22:07:34 2008 Host : Das Ermes # : 260132919 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1728 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Rail Thermal Sensor Failure (OPEN) Sensor: Rail Center Channel

Error: 260132914 Rail Temperature Out Of Range See Long Description for complete error Temperatures vary. Wed Jun 21 14:51:53 2006 Error: 260132914 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2071 Wed Jun 21 14:51:52 2006 Host : Das Ermes # : 260132914 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2071 Function : No System Function Reported Scan Type : None/Unknown Scan: 81/201/0 Rail Temperature Out Of Range Sensor: Rail Low Channel Temperature: 3640 <--- In this case, temp was always too LOW Average A/D Temperature: 0 Limits: 3650, 3950

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5129987 Low channel thermostat asm for VCT detector See long description for details Part #: 5129987 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Low channel thermostat asm FRU: Yes Status: Orderable Comments: **FE ASSIST** P/N 5129987 replaced by 5261028 RoHS (212001261 June 16,2010) 5124291 cable, das routing board, Used on VCT see long description for details Part #: 5124291 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: cable, das routing board, vct FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues