Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT244540244634260134314

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state

Error: 244634 Subsystem State; DCB:down, CCB:down, TCB:down, JEDI:down, TGP/ORP:down, All Down See Long Description for complete error Thu Mar 18 21:29:36 2010 Error: 244634 Host: sim1 Process: fwmgr File: fwmgr utils.c Line: 664 Running: System Reset Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:down TCB:down JEDI:down TGP/ORP:down

Error: 260134314 CAN Watchdog response timeout - Stopping drive motion See Long Description for details Error: 260134314 Host: tgp Process: tgp_subsys File: gtcb_can_wdog.cxx Line: 770 Wed Feb 13 03:16:13 2008 Host : Table Sub-System Ermes # : 260134314 Exception Class : Secondary Severity : Pri/Soft File : gtcb_can_wdog.cxx Function : No System Func

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5197802 Cherokee 5kW 48vdc power supply with fuse board. USED ON VCT gantry rotating frame. Supercedes previous Pioneer part numbers 5197802, 5183361, 5154556, and 5126259 Part #: 5197802 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCTHD 48V PS with FRU packaging FRU: Yes Status: Orderable Comments: **FE ASSIST** 5197802-2 is used on a HD-CT750 system, Do not use on a VCT system.(212001261 9/30/2010) .Per HDVCT RSPL, 212008898, 11/6/08 Part #: 5197802 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Pioneer 5kW Power Supply Assembly Collector, with FRU packaging material FRU: Yes Status: Orderable Comments: Per ECR 2037621, 212008898, 1/5/07. If no stock, please order part # 5183361 --Description: VCT 5kW Power Supply Assembly, with ferrite and external shield. Part number change up history information: P/N 5197802 Cherokee " supply has two main differences: 1) The newer Cherokee supply is approximately 1.2 kg (2.6 lbs) Heavier then the older Pioneer Supply - this may cause a Gantry rebalance. 2) The large Ferrite bead is not visible on the Cherokee Supply because the ferrite is internal on the Cherokee Supply. P/N 5183361 was created to add an external shield that diverts the magnetic field energy away from the low channel DAS/Detector channels. The low channels on the VCT DAS/Detector are sensitive to magnetic fields that are produced from the power supply. This p/n replaced 5154556. P/N 5154556 was created to add a ferrite core to the 48 VDC output cable of the supply. This change was made to address streaking imaging artifacts that were sometimes seen with low mA scans by decreasing noise induced into the DAS/Detector. This p/n replaced 5126259. P/N 5126259 was the original VCT gantry rotating frame 48 vdc power supply. 5183361 48 vdc power supply with ferrite core and external shield 5154556 48 vdc power supply with ferrite core 5126259 48 vdc power supply without ferrite core or magnetic shield
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.