Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18310833003

LightSpeed 2.X (Plus, QX/i) - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183108 KV fault detected: Cathode shoot-thru. See long description for full error Sun Jun 5 23:17:05 2005 Error: 183108 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 248 Function: XRAY GENERATION : KV Loop Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:05:545 Log Series: 424 KV fault detected: Cathode shoot-thru. Fault latch address = ffcff3 hex, Bit = D3

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2208852 - HP INVERTER RHAPSODE - used on CT/i, 1.x, 2.x, 3.x and 4.x systems. 2208852 HP INVERTER RHAPSODE 2208852 HP INVERTER RHAPSODE Part #: 2208852 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HP INVERTER RHAPSODE. FRU: Yes Status: Orderable Comments: IF RHAPSODE SYSTEM CURRENTLY HAS T3202JK CAN USE 2208852 . THE LONGER BOLTS NEEDED FOR THE 2208852 INVERTER ARE NORMALLY SUPPLIED IN A BAG WITH THE INVERTER OR CAN BE ORDERED SEPARATELY UNDER P/N 46-208561P43
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.