GE Healthcare · Nuclear MedicineInfinia
Acquisition halts Error #47 PSD test failure Hardware failure
Symptom
Acquisition halts Error #47 PSD test failure Hardware failure Acquisition halts, Error #47 PSD test failure Hardware failure
Error message
Error # 47 (Error 47) ACQUISITION HALTS PSD TEST FAILURE Corrective Action: Replace the Collimators. If the problem persists, replace the Collimator ID board on both collimators Check Connectors P7 and P405/P1205 on HCIC board of Head #1 and Head #2, respectively. Check interior Head cable (CBL001314) between HCIC P7 and the CBCIC J4 for both Heads. Check the cnnection of Connector J1 on the CBCIC board for both Heads. Replace the FOXPOP board.
Applies to
- Models
- Infinia
- Product
- Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System
- Component
- 5338523 Two Collimator ID Boards (Kit) 2199382 Collimator ID Board Part replaced by 5338523 . This part is for the new Collimator ID design. Please refer to SN 43837 for instructions If your system is equiped with the old design (2199382) then you will need to order 5342601 - First Time Collimator ID UPgrade Kit ( FMI 40822 ). Note: The screws holding the ID pcb's on the collimator may have been secured with hard or permanent loctite making it very hard to remove the screws The screws that holding the Collimator ID board is defined to be secured by Loctite 222 (that can be untightened later). To open the screws should use the right tool - Phillips screwdriver PH #2. *Collimator/ Collimator cart Collimator/ Collimator cart
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.