Acquisition halts Error #47 PSD test failure Hardware failure

Symptom

Acquisition halts Error #47 PSD test failure Hardware failure Acquisition halts, Error #47 PSD test failure Hardware failure

Error message

Error # 47 (Error 47) ACQUISITION HALTS PSD TEST FAILURE Corrective Action: Replace the Collimators. If the problem persists, replace the Collimator ID board on both collimators Check Connectors P7 and P405/P1205 on HCIC board of Head #1 and Head #2, respectively. Check interior Head cable (CBL001314) between HCIC P7 and the CBCIC J4 for both Heads. Check the cnnection of Connector J1 on the CBCIC board for both Heads. Replace the FOXPOP board.

Applies to

Models
Infinia
Product
Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System
Component
5338523 Two Collimator ID Boards (Kit) 2199382 Collimator ID Board Part replaced by 5338523 . This part is for the new Collimator ID design. Please refer to SN 43837 for instructions If your system is equiped with the old design (2199382) then you will need to order 5342601 - First Time Collimator ID UPgrade Kit ( FMI 40822 ). Note: The screws holding the ID pcb's on the collimator may have been secured with hard or permanent loctite making it very hard to remove the screws The screws that holding the Collimator ID board is defined to be secured by Loctite 222 (that can be untightened later). To open the screws should use the right tool - Phillips screwdriver PH #2. *Collimator/ Collimator cart Collimator/ Collimator cart
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.