LightSpeed 5.X (60kW)
Symptom
Scanner fails to boot to applications, Unable to verify DARC system startup. The system will not boot up to applications with a pink attention box: Unable to verify system functionality on startup. Please power cycle the console. If the system fails to startup after the power cycle, then please call GE Service. FE or OLE sees that they can telnet to the DARC but they cannot ping or rsh with the DARC. If the DARC is observed while booting, Drive X adapter X predicting future failure Press any key to continue is seen.
Error message
Error: 260006806 Axial Motor Control (AMC) logic will NOT accept commands without a Valid Configuration/Characterization file. See the Long Description for the full error code Thu Nov 18 20:34:37 2004 Error: 260006806 Host: STC Process: Axial File: amc_exec.c Line: 3464 Function: No System Function Reported Scan: 50004/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 20:34:37:390 Log Series: 37 Axial Motor Control (AMC) logic will NOT accept commands without a Valid Configuration/Characterization file. (Cmd: 0x1804ffff). Do Not Use - use one of the other Error: 290014004 Unable to verify DARC system startup See Long description for complete error NOTE: This is error does not mean you need to replace the darc. Troubleshooting is necessary This error comes in different contexts per the below Mon Dec 12 08:00:11 2005 Error: 290014004 Host: ct99 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 172 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for ping darc response (exit status: 8) Fri Dec 8 14:42:12 2006 Error: 290014004 Host: kghcateqxi Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS See LD for more info 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS Part #: 5114772 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk H16 MDAS, GDAS FRU: Yes Comments: Added per GOC4 RSPL Replace the DARC2: 1) If you get a Westville DARC P/N 5114772 load the DARC software to recover. 2) If you get Jarrell p/n 5147442 perform LFC using 06MW29.7 software.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.