Pular para o conteúdo principal
GE Healthcare · CTHiSpeed184514

Scanner Hardware Stopped Scan message during axial scan

Symptom

Scanner Hardware Stopped Scan message during axial scan. Long Description for Feedback Only. Scanner Hardware Stopped Scan message during axial scan

Error message

Error: 184514 HW Error: Possible PDU (servo amplifier) or axial board problem Please see the long description for the full error code Tue Jun 3 21:13:05 1997 Suite: VACT Host: STC Proc: Control Error: 184514 File: resetServoAmp.c 1.8 Method: No Method Line: 87 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 1031 Log Series: 20 Could not clear servo amplifier fault. HW Error: Possible PDU (servo amplifier) or axial board problem

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-186852P2 A1K2 contactor See long description for details and picture Part #: 46-186852P2 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: RELAY (K2 BACK-UP CONN.) FRU: Yes Status: Orderable Comments: FLS - ALSO AVAILABLE AS 46-186852P2UUSE UP THAT STOCK FIRST! SWITCH ON SIDE OF RELAY= 46-296690P1 A4K2 PF8 TWICE FOR FE ASSIST COMMENTS!!! FOR FE ASSIST PF8 AGAIN!!! FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PARTS WHEN AVAILABLE., IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. FLS - FE ASSIST MS6588 10/2000 46-186852P2 = AXIAL RELAY ARE YOU ORDERING THE CORRECT PART?? FOR CT PDU THE AXIAL RELAY ASMIS MADE UP OF TWO RELAYS, THE 46-186852P2 & 46-296690P1 P1 RELAY MOUNTED ATOP OF P2. 46-186852P2 = AXIAL LOOP CONTACTOR RELAY. 46-296690P1= AUXILIARY CONTAC T RELAY. SEE CTI FRU MANUAL PAGE 5-5 REVISION 26 AND ABOVE, OLDER REV DO NOT SHOW THE AUX RELAY. DO YOU STILL WHAT TO ORDER THIS PART??
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.