GE Healthcare · CTHiSpeed185122

Replace 24 VDC table PS

Symptom

Table will not move. Table will not move

Error message

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296316P2 TABLE POWER SUPPLY See Long Description for Detail Part #: 46-296316P2 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: TABLE POWER SUPPLY FRU: Yes Status: Orderable Comments: FLS - 46-296316P1 REPLACED BY 46-296316P2 PER PCN 189228. 46-296316P2 IS BACKWARDS COMPATIBLE TO P1 PER WES HEIMAN. MS4636 4/5/95 24VDC @ 7.5A OUTPUT POWER SUPPLY. USED ON ZEUS TABLE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.