Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18288516417182884

Operator reported that table foot pedals had stopped working

Symptom

Table foot pedals malfunction. Table foot pedal malfunction.

Error message

Error: 182885 The Home Foot Pedal Is STUCK. See LD for sample error message. Mon Oct 17 20:14:51 2005 Error: 182885 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 631 Function: No System Function Reported Scan: 16417/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:14:51:975 Log Series: 241 The Home Foot Pedal Is STUCK.

Error: 182884 The Load Foot Pedal Is STUCK. See LD for full sample error message. Mon Oct 17 20:14:50 2005 Error: 182884 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 623 Function: No System Function Reported Scan: 16417/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:14:50:925 Log Series: 240 The Load Foot Pedal Is STUCK.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296129G1 FOOTSWITCH used on Zeus style tables 1x through 4x plus CTi and HSA See long description for details Part #: 46-296129G1 Part Type: GP CONSUMABLES Part Weight: 3.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FOOTSWITCH FRU: Yes Status: Orderable FLS - THIS IS A 9-PIN MOLEX CONNECTOR USED WITH THE CT/I HISPEED, QX AND QXI PLUS CT SYSTEMS. THIS MOVED TO GEHW AS PART OF THE GPC CHINA TABLE MOVE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues