Pular para o conteúdo principal
GE Healthcare · CT

Sytec 2000/3000/4000

Symptom

Unable to scan with XG errors

Error message

ER-117 XG Fail - contains information on all XG ER117 errors See Long Description ER-117 XG Fail ER–117 ER – 117 : XG Error 00000XYY Error Code X = _ 0: Test Mode 1: Stand–by 2: Preparation Acting 3: Generator Ready 5: Exposure 7: Transition Step to Stand–by ER–117 (continued) ER – 117 : XG Error 00000XYY Error Code YY = _ X01: No DC550V X02: 15V_Safe Error X03: Over Voltage Capacity X11: Rotor Controller Error X12: Rotor Acceleration Error X13: Rotor Feedback Test X14: Rotor Fault Detect Test X16: Low Current Error X18: Unbalance Error X21: Over Filament Current 021: Feedback Test X22: No Filament Current 022: Fault Detect Test X23: Bad Relay Return 023: Small Focus Test X24: No Filament Voltage 024: Large Focus Test X25: Filament Control Error 526: MA measure Error 031: Feedback Test 032: Fault Detector Test 033: Close Loop Test 531: MA Accuracy Error 532: KV Accuracy Error 533: Spit Error 534: Exposure Backup Timer Error 535: Anode Over mA 536: Cathode Over mA 537: Anode Over kV 538: Cathode Over kV 539: Anode Shoot Through 540: Cathode Shoot Through 541: Anode Over Current 542: Cathode Over Current X43: HV ON with Error 544: KV Unbalanced Error X48: Generator Safety Loop Open X

Applies to

Product
Sytec 2000/3000/4000 and "i" Series and, Vertex PreSage, Luminex Vj Lj SLj Sytec 2000/3000/4000 and "i" Series and, Vertex PreSage, Luminex Vj Lj SLj Sytec 2000/3000/4000 and "i" Series and, Vertex PreSage, Luminex Vj Lj SLj Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3301T / D3302T Xray tube used on Sytec 2000/3000/4000 and "i" Series, Vertex PreSage, Luminex Vj Lj SLj, Pace, and Pace Plus See long description for more information. D3302T Xray tube used on Sytec 2000/3000/4000 and "i" Series, Vertex PreSage, Luminex Vj Lj SLj, Pace, and Pace Plus Part #: D3302T Part Type: GP TUBES Part Weight: 84.10 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE SYTEC AND SYTEC I FRU: Yes Status: OrderableComments: FLS - NEW 1994 TUBE CAT. EXCHANGE. REPLACES D1193GR .9 X .7 FS SYTEC 2000, 3000, 4000 OR PACEAND I SERIES (GREEN TUBE PART NUMBER IS: D3301T) DIMS 27X37X28H (INCHES) *MAXIRAY VERSION OF THIS TUBE IS D3309T* AM SOURCE = MEXICO FLS - FE ASSIST MS6588 FEB 2002 D3302T = GLASS TUBE YMS. DUE TO HIGH LEVEL OF DOA TUBESPLEASE REFER TO THE FOLLOWING SERVICE NOTES SN 108, T8011 & IF YOU HAVE A KYOSAN GENERATORSEE T6005 TO HELP ISOLATE THE ISSUE. ALSO IF YOU HAVE ANY DOUBTS PLEASE ASK YOUR LOACL ZSE OR CALL THE ONLINE CENTER. MAXIRAY EQUIVALENT IS D3309T
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.