Pular para o conteúdo principal
GE Healthcare · CT24201422955

Lightspeed 1.X - Octane

Symptom

Recon Status Indicates: Recon Resetting; Intermittent recon retries; Image recon delay noticed by cust See Long Description. Not for GRE Recon Status Indicates: Recon Resetting; Intermittent recon retries; Image recon delay could sometimes be noticed by customer. Issue could also cause SmartPrep exams to seem to hang. IG DSP encountered scan data view checksum error and the iceConsole.log posted repeated RECON_IG_RESET messages. Issue and errors can be very intermittent.

Error message

Error: 242014 DSP encountered scan data view checksum error on view num Host: IG Process: DSP Function: Image Generation : Back Projection See LD for complete error Wed Nov 17 08:40:20 2004 Error: 242014 Host: IG Process: DSP File: prepig.c Line: 542 Function: Image Generation : Back Projection Job Type: Prospective Helical Scan: 22955/3/1 Image: 22955/4/44 Recon Seq Id: 552

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2118899-2 RHAPSODE IMAGE GENERATOR Board (IG Board). Used in Lightspeed 1x, CTi 4x, CTi 6x systems. See LD for details Part #: 2118899-2 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: RHAPSODE IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - IMAGE GENERATOR BOARD USED ON RHAPSODE MS1356 BURR 12-95 REPLACED 2118899 PER PCN 189598. MS4636 2/6/96 PCN 189939 GELS - KANBAN CSS F00,PCB IMAGE GENERATOR
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.