Symptom

Scanner Hardware Stopped Scan with Tube casing temperature has exceeded 92 degree - document in RFS - critical issue Scanner Hardware Stopped Scan Tube casing temperature has exceeded 92 degree C.

Error message

Error: 260118275 (110-0903) Tube casing temperature has exceeded 92 degree C See LD for more details Sun Nov 6 12:44:33 2005 Error: 260118275 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sun Nov 6 12:44:32 2005 Host : Jedi Ermes # : 260118275 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50108/1/3 (110-0903)Tube casing temperature has exceeded 92 degree C.^C Z6,Z7= 0h,0h

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2370625 Hercules Pump/Adapter/Hose. 2370625 Hercules Pump/Adapter/Hose. Part #: 2370625 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HERCULES PUMP/ADAPTER/HOSE ASM. FRU: Yes Status: Orderable Comments: FLS - HERCULES PUMP/ADAPTER/HOSE USED ON: ON HPOWER ASM. MS13452 031031 RISK: SITE DOWN FREQUENCY: LOW PDR ON FILE 8413.70.2040 IM05 REQUESTED
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.